Performance requirements and goals for the MIRI imaging are:
1.) Imaging shall extend from 5 to 27 microns in at least 12 spectral bands
2.) Images shall be diffraction limited at 5 microns and shall be critically sampled at 6(TBR) microns
3.) The imaging field of view shall be at least equivalent to 1.5 x 1.5 arcmin
4.) The imager shall be natural background limited between 5 and 10 microns
5.) The imager shall be able to measure point sources at least as bright as 4mJy at 10 microns
6.) The imager shall include a low resolution spectroscopic capability (R ~ 100) from 5 to 10 microns
7.) 'Simple' coronagraphs shall be provided, simple in the sense that they will not require additional mechanisms.
Performance requirements and goals for the MIRI spectroscopy are:
1.) Spectroscopy shall be provided between 5 and 27 microns, with extension to 28.3 microns as a goal
2.) The nominal spectroscopic resolution shall be ~3000 (or more) from 5 to 10 microns, and shall be reduced by diffraction to no less than 1400 at 20 microns
3.) The spectrometer shall include integral field units with fields of view of 2 arcsec, with fields of view of 3 arcsec as a goal
4.) The nominal pixel size in the spectrometer shall be 0.2 arcsec
These requirements were set by an international team (The Mid-Infrared Steering Committee) in July, 2001, and have formed the basis for the subsequent design definition.